Nitric oxide exhaled breath analysis
The Sievers NOA-280i nitric oxide analyser uses a patented* technique and
the On-Line NO Breath Kit to measure on-line concentrations of NO in exhaled breath without nasal contamination. The built-in exhaled pressure feedback system and optional Thermal Mass Flowmeter
facilitate exhalation at constant flow rates from 10–1000 mL/s. The NOAnalysis REB PC software is used for flow meter calibration or restrictor selection, data collection and processing. For subjects who cannot perform the on-line or off-line techniques. (i.e., ventilated subjects, small children, animals), a tidal breathing software
package is used that determines the beginning and end of each exhalation and the concentration of NO for each breath.For off-line measurements there are patented Bag Collection and Sampling Kits that permit collection of exhaled breath for later analysis. Two kits are available: Vital Capacity Kit for
collection of exhaled breath at 350 mL/s (ATS recommendation) and the Deadspace Discard Kit that permits collection of exhaled breath under the same conditions as on-line measurements. The concentration of NO in the
Mylar® bags is stable for 8 to 12 hours and the bags can be reused after a simple cleaning procedure. The NOAnalysis Bag program is used for analysis of the samples.