Milestone mercury analyser
The DMA-80 Mercury Analyser effortlessly analyses liquid and solid samples without sample
digestion or wet chemistry pre-treatment steps, and without any waste disposal. Sample analysis is performed in 5 minutes and up to 40 samples can be automatically processed in less than 4 hours. An intuitive
controller uploads sample weights, controls the analysis and processes data with built-in report generation and networking capabilities
Liquid or solid samples are introduced into the DMA-80 Mercury
Analyser autosampler. Prior to combustion, the sample is initially dried in an oxygen stream passing through a quartz tube located inside a controlled heating coil. The combustion gases are further decomposed on a
catalytic column at 750 °C. Mercury vapour is collected on a gold amalgamation trap and subsequently desorbed for quantitation. Mercury content is determined using atomic absorption spectrometry at 254 nm, and the
results are displayed using a Pentium-based control terminal
The large touch-screen interface
provides easy, graphical interaction and allows manual operation of the instrument, as well as pre-programmed, unattended automatic operation. Parameters can be changed in the middle of a sample run and there is no
need to wait for an entire queue of samples to be processed.
A typical peak profile resulting from analysis is shown where the 1st peak corresponds to the long pathlength flow cell and Hg content up to 20 ng. The
instrument determines the absolute amount of Hg (normally expressed in ng), and then the software calculates its concentration in the sample.
The 2nd peak corresponds to the short pathlength flow cell and Hg
content in the range of 20-1000 ng..The software measures the height, the half width and the area of the absorbance peak, that assists in both method development and troubleshooting. Data can be easily exported into
a spreadsheet.
Secure electronic data keeping made easy with Milestone's software
that fully complies with CFR 21 part 11
- Easily identifiable data files
- Automatic storage of data
- Editable process parameters for individual samples
Intuitive EasyControl software
Control terminals and data analysis